Cookies på Omegas websteder Vi anvender cookies på vores websted. Disse cookies er afgørende for, at webstedet kan fungere korrekt. Hvis du fortsætter uden at ændre dine indstillinger, antager vi, at du gerne vil modtage alle de cookies, der anvendes på dette websted. Du finder flere oplysninger om cookies ved at klikke her
The TX206TH is a cost effective "smart" in head transmitter that accepts thermistor temperature sensors and converts sensor output over a configured range to a standard industrial (4 to 20) mA transmission signal.
A variety of thermistor types are available when using the configuration software: select thermistor type, temperature range, units and burnout direction, without requiring calibration equipment.
PC configuration is performed quickly using our new USB port driven configurator by simply connecting two clips to the TX206TH loop terminals and following the software instructions. Calibration set up may be saved as a file on the PC for later use.
Alternatively, a single push button and LED indicator allows the user to configure transmitter range against a manually set input condition. A red LED is included to help guide the user. The LED also operates as a sensor error indicator.
Transmitters can be factory configured for any range specified at the time of order, removing the need for user configuration.
Specifications (at 20°C)
INPUT
Sensor Type: Thermistor Sensor Connection: Screw terminal Minimum span: 5 °C Accuracy (typical): ± 0.2 °C ±0.2 % of reading Thermal Drift: ± 0.02 % / °C
OUTPUT
Output Type: 2 wire (4 to 20) mA current loop Output range: (4.0 to 20.0) mA Output Connection: Screw Terminal Maximum output: 21.5 mA Minimum output: <3.75 mA Accuracy: (mA output /2000) or 5 µA (Whichever is the greater) Loop Voltage effect: 0.2 µA / V Thermal drift: ±2 µA / °C Maximum output load: [(Vsupply-10)/21]K Ω;(Example: 666 Ω @ 24 V)
THERMISTOR TYPES
Please refer to configuration software for complete up to date list:
3KB (44005, 44030), 5KB (44007, 44034), 10KB (44016,44036)
10KH (44006, 44031), 30KH (44008), 2252KB (44004, 44033)
GENERAL SPECIFICATION
Response time: 1 second Start-up time: 2 seconds ( I out < 4 mA during start up) Warm-up time: 1 minute to full accuracy Power Supply: (10 to 30) Volts dc
ENVIRONMENTAL
Ambient operating range: (-40 to +85) °C Ambient storage temperature: (-50 to +90) °C Ambient humidity range: (10 to 90) % RH non condensing
PHYSICAL
Dimensions: 43 mm diameter; 21 mm height Mounting: Two 5.5 mm Ø holes on 33 mm centres, will fit a DIN standard termination head. Weight: 31 g (encapsulated)